MeasurementsMXO 4 Series
230User Manual 1335.5337.02 ─ 07
9.2.4.4 Counting
Meas. type Symbol Description/result
Pulse count PCount The number of positive or negative pulses of the
waveform, or of both positive and negative pulses.
The mean value of the signal is determined. If the sig-
nal passes the mean value, an edge is counted. A
positive pulse is counted if a rising edge and a follow-
ing falling edge are detected. A negative pulse is
counted if a falling edge and a following rising edge
are detected.
N321
N321
Edge count EdCo The number of positive or negative edges, or of both
positive and negative edges.
The instrument determines the mean value of the sig-
nal and counts an edge every time the signal passes
the mean value.
N321
N321
9.2.5 Settings for measurements
In the "Measurement" dialog, you can enable measurement and define their settings.
9.2.5.1 Setup tab
Access: "Menu" > "Measurement" > "Setup" tab.
Automatic measurements