TriggerMXO 4 Series
163User Manual 1335.5337.02 ─ 07
"Positive"
Rising edge, a positive voltage change.
"Negative"
Falling edge, a negative voltage change.
"Both"
Both the rising and the falling edge.
Remote command:
TRIGger:EVENt<ev>:SETHold:CSOurce:EDGE on page 560
Clock level
Sets the voltage level for the clock signal.
Both the clock level and the clock edge define the starting point for calculation of the
setup and hold time.
Remote command:
TRIGger:EVENt<ev>:SETHold:CSOurce:LEVel on page 560
Setup time
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
Remote command:
TRIGger:EVENt<ev>:SETHold:STIMe on page 561
Hold time
Sets the minimum time after the clock edge while the data signal must stay steady
above or below the data level.
Remote command:
TRIGger:EVENt<ev>:SETHold:HTIMe on page 561
7.4.11 State trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = State"
The state trigger verifies if the channel states match the defined pattern at the clock
edge. The trigger occurs if the logical combination of the input channels is true at the
crossing point of the selected clock edge and the trigger level.
This section describes the trigger-type specific settings. For settings that apply to all
trigger types, see Chapter 7.2, "Common trigger settings", on page 142.
Trigger types