TRIGger:MEVents:SEQuence<se>:RESet:TIMeout:TIME <ResetTimeout>
The time the instrument waits for the number of B-events specified using TRIGger:
MEVents:SEQuence<se>:COUNt, before the sequence is restarted with the A-trig-
ger.
Suffix:
<se>
.
Irrelevant, omit the suffix.
Parameters:
<ResetTimeout> Range: 0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Enable reset by time, Reset timeout" on page 146
TRIGger:MEVents:SEQuence<se>:RESet:TIMeout[:ENABle]
<State>
If set to ON, the instrument waits for the time defined using TRIGger:MEVents:
SEQuence<se>:RESet:TIMeout:TIME for the specified number of B-events. If no
trigger occurs during that time, the sequence is restarted with the A-event.
Suffix:
<se>
.
Irrelevant, omit the suffix.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Enable reset by time, Reset timeout" on page 146
17.9.3 Edge trigger
TRIGger:EVENt<ev>:EDGE:SLOPe.................................................................................544
TRIGger:ANEDge:LEVel................................................................................................. 545
TRIGger:ANEDge:COUPling........................................................................................... 545
TRIGger:ANEDge:FILTer................................................................................................ 545
TRIGger:ANEDge:CUToff:HIGHpass................................................................................546
TRIGger:ANEDge:CUToff:LOWPass................................................................................ 546
TRIGger:ANEDge:NREJect.............................................................................................546
TRIGger:EVENt<ev>:EDGE:SLOPe <Slope>
Sets the edge direction for the trigger.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Trigger