Mixed signal option (MSO, R&S MXO4-B1)MXO 4 Series
425User Manual 1335.5337.02 ─ 07
"Robust"
Sets different hysteresis values for falling and rising edges to avoid
an undefined state of the trigger system. Use this setting for very
noisy signals.
Remote command:
PBUS<pb>:HYSTeresis<n> on page 851
Active logic channel
Selects the state of the respective bus channel.
Enable/Disable D7-D0,D15-D8
The buttons select or deselect all digital channels of a pod at once.
14.1.2 Bus
Access: "Menu" > "Logic" > "Bus" tab
Show bus
If enabled, the resulting bus signal and bus values are displayed in the diagram.
Remote command:
PBUS<pb>:DISPlay:SHBU on page 850
Data format
Sets the data format of bus values, which are displayed in the decode table and on the
comb bus display.
Available formats are: Hex, octal, binary, ASCII, string, signed, unsigned, symbolic, and
auto.
Signed and Unsigned are integer data types with a maximum 16-bit length. Unsigned
is used for positive integers. Signed is used for positive and negative integers.
Logic configuration