Protocol analysisMXO 4 Series
398User Manual 1335.5337.02 ─ 07
Condition
Sets the operator to define a specific data pattern ("Equal" or "Not
equal") or a range smaller or bigger than a specific pattern.
"Pattern"
Defines the data pattern.
"Position"
Sets the number of data events that are ignored, before trigger condi-
tion check of the data starts.
Remote command:
TRIGger:LIN:DCONdition on page 799
TRIGger:LIN:DMIN on page 799
TRIGger:LIN:DPOSition on page 799
Error conditions
Triggers if one or more of the following errors occur.
"Parity error"
Parity bits are the bits 6 and 7 of the identifier. They verify the correct
transmission of the identifier.
"Sync error"
Synchronization error.
"Checksum
error"
The checksum verifies the correct data transmission. It is the last byte
of the frame response. The checksum includes not only the data but
also the protected identifier (PID).
Remote command:
TRIGger:LIN:IPERror on page 800
TRIGger:LIN:SYERror on page 801
TRIGger:LIN:CHKSerror on page 799