Waveform generator (option R&S MXO4-B6)MXO 4 Series
438User Manual 1335.5337.02 ─ 07
Pulse width
Sets the pulse width, the pulse duration of the generated pulse waveform.
Remote command:
WGENerator<wg>:FUNCtion:PULSe[:WIDTh] on page 858
Default setup
Presets the generator to a default setup. The default includes the following settings:
"Function type" = "Sine"
"Frequency" = "1 MHz"
"Amplitude" = "1 Vpp"
Remote command:
WGENerator<wg>:PRESet on page 860
DC level
Available for "Function type" = DC.
Sets the level for the DC signal.
Remote command:
WGENerator<wg>:VOLTage:DCLevel on page 860
Noise level for DC
Available for "Function type" = "DC".
Sets the level of the noise for the DC signal.
Remote command:
WGENerator<wg>:MODulation:NDCLevel on page 867
Period
Available for "Function type" = "Pulse".
Sets the period of the pulse waveform.
Remote command:
WGENerator<wg>:PERiod on page 859
15.1.2 Modulation settings
Modulation is available only for sine and square waveforms.
Access: "Gen" signal activator > "Gen"1 or "Gen"2 > "Setup" tab > "Function type" =
"Sine" or "Square" > "Modulation".