Remote control commandsMXO 4 Series
860User Manual 1335.5337.02 ─ 07
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Period> Range: 1E-08 to 1000
Increment: 1
*RST: 1E-06
Default unit: s
Usage: Asynchronous command
Manual operation: See "Period" on page 438
WGENerator<wg>:PRESet
Presets the generator to a default setup. The default includes the following settings:
"Function type" = "Sine"
"Frequency" = "1 MHz"
"Amplitude" = "1 Vpp"
Suffix:
<wg>
.
1...2, index of the waveform generator
Usage: Setting only
Asynchronous command
Manual operation: See "Default setup" on page 438
WGENerator<wg>:VOLTage:DCLevel <DCLevel>
Sets the level for the DC signal, if WGENerator<wg>:FUNCtion[:SELect] is set to
DC.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<DCLevel> Range: -5 to 5
Increment: 0.01
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "DC level" on page 438
WGENerator<wg>:VOLTage:HIGH <High>
Sets the high signal level of the output waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Waveform generator (option R&S MXO4-B6)