Remote control commandsMXO 4 Series
859User Manual 1335.5337.02 ─ 07
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<FunctionType> SINusoid | SQUare | RAMP | DC | PULSe | SINC | CARDiac |
GAUSs | LORNtz | EXPRise | EXPFall | ARBitrary
SINC: Cardinal sine
*RST: SINusoid
Usage: Asynchronous command
Manual operation: See "Function type" on page 431
WGENerator<wg>:FUNCtion[:SQUare]:DCYCle <SquareDutyCycle>
Sets the duty cycle for the pulse function.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<SquareDutyCycle> Range: 0.01 to 99.99
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Duty cycle" on page 437
WGENerator<wg>:OUTPut[:LOAD] <Load>
Select the user load, the load of the DUT at its connection.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Load> FIFTy | HIZ
FIFTy: 50Ω
HIZ: High-Z (high input impedance)
*RST: HIZ
Usage: Asynchronous command
Manual operation: See "User load" on page 437
WGENerator<wg>:PERiod <Period>
Sets the period of the pulse waveform, if WGENerator<wg>:FUNCtion[:SELect] is
set to PULSe.
Waveform generator (option R&S
MXO4-B6)

Table of Contents