Instrument setupMXO 4 Series
97User Manual 1335.5337.02 ─ 07
As long as no user-defined preset file is loaded and "Enable user-defined preset" is
disabled, the [Preset] key also resets the instrument settings to factory defaults.
5.7 Maintenance settings
In the "Settings" > "Maintenance" dialog box, you can update the firmware, perform
self-alignment.
5.7.1 Firmware update
Access: "Menu" > "Settings" >"Maintenance" > "FW update".
Your instrument is delivered with the latest firmware version. Firmware updates are
provided on the internet at:
www.rohde-schwarz.com/firmware/mxo4.
The "Release Notes" describe the improvements and modifications of all firmware ver-
sions. They also explain how to update the firmware. They are available along with the
firmware on the same web page.
Remote commands:
SYSTem:FW:FILepath on page 486
SYSTem:FW:STARt on page 487
5.7.2 Alignment
When data from several input channels is displayed at the same time, it can be neces-
sary to align the data. Alignment is done vertically or horizontally to synchronize the
time bases or amplitudes and positions. Perform an alignment, for example, when
strong temperature changes occur (> 5°).
5.7.2.1 Alignment settings
Access: "Menu" > "Settings" > "Maintenance" > "Alignment"