Remote control commandsMXO 4 Series
548User Manual 1335.5337.02 ─ 07
17.9.5 Width trigger
TRIGger:EVENt<ev>:WIDTh:DELTa.................................................................................548
TRIGger:EVENt<ev>:WIDTh:POLarity..............................................................................548
TRIGger:EVENt<ev>:WIDTh:RANGe...............................................................................548
TRIGger:EVENt<ev>:WIDTh:WIDTh................................................................................ 549
TRIGger:EVENt<ev>:WIDTh:DELTa <WidthDelta>
Defines a range around the width value.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 5E-10
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "±Delta" on page 152
TRIGger:EVENt<ev>:WIDTh:POLarity <Polarity>
Sets the polarity of a pulse, which is the direction of the first pulse slope.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Polarity> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Polarity" on page 152
TRIGger:EVENt<ev>:WIDTh:RANGe <RangeMode>
Selects how the range of a pulse width is defined.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
*RST: WITHin
Usage: Asynchronous command
Manual operation: See "Range" on page 152