Column Description
CRC Frame Check (Cyclic Redundancy Code, CRC)
Bit rate Value of the bit rate
Table 13-14: Content of the Details frame table
Column Description
Label Label of the field
State State of the field
Value Value of the field
Data format
You can select various data formats for the values displayed in the decode table and in
the combs of the decoded signal.
Remote command:
ā
SBUS<sb>:FORMat on page 704
Export of decode results
1. In the protocol "Shortcuts" tab, press "Export results".
The "Export results" dialog opens.
For details, see Chapter 13.1.7, "Export protocol results", on page 318.
2. Select the details that you want to export and the file format.
3. Tap "Save as".
4. Key in a name and select the file format.
Remote commands
Remote commands to retrieve decode results are described in Chapter 17.16.8.3,
"Decode results", on page 835.
13.8.4 Performing 10BASE-T1S decoding
This chapter explains step by step how to configure and decode the 10BASE-T1S bus.
13.8.4.1 Configuring 10BASE-T1S signals
For configuration, assign the lines to the input channels, and define the active states
and the logical thresholds.
For details on configuration settings, see Chapter 13.8.1, "10BASE-T1S configuration",
on page 414.
1. Tap the "Bus" activator in the bottom right of the screen.
10BASE-T1S (option R&S
MXO4-K560)