Mixed signal option (MSO, R&S MXO4-B1)MXO 4 Series
426User Manual 1335.5337.02 ─ 07
If the target file format is BIN, you can save only signed and unsigned binary data. The
data format "Signed" writes signed data; all other formats are saved as unsigned
binary data.
Remote command:
PBUS<pb>:DATA:FORMat on page 855
PBUS<pb>:DATA:HEADer? on page 856
PBUS<pb>:DATA[:VALues]? on page 856
Clocked bus
If a bus is a clocked bus, one of the digital channels serves as the clock of the bus.
For an unclocked bus, the logical state of the bus is determined for each sample. For a
clocked bus, the logical state is determined only at the specified clock edges.
Bus clocked ← Clocked bus
Enable this option for a clocked bus.
Remote command:
PBUS<pb>:CLON on page 848
Show decode table ← Clocked bus
If enabled, a result table is shown with decoded values and corresponding points in
time of the bus signal. Each clock edge corresponds to one row in the table.
The decode table is only available for clocked buses to check the data words.
Remote command:
PBUS<pb>:DECTable:SHOW on page 849
PBUS<pb>:DECTable:COUNt? on page 849
PBUS<pb>:DECTable:DATA? on page 850
Clock source ← Clocked bus
Selects the digital channel used as clock.
Remote command:
PBUS<pb>:CLOCk on page 848
Clock slope ← Clocked bus
Selects the slope of the clock signal at which all digital channels of the bus are ana-
lyzed.
Remote command:
PBUS<pb>:CLSLope on page 849
14.1.3 Label settings
Access: "Menu" > "Logic" > "Label" tab
Logic configuration