Remote control commandsMXO 4 Series
849User Manual 1335.5337.02 ─ 07
Manual operation: See "Bus clocked" on page 426
PBUS<pb>:CLSLope <ClockSlope>
Selects the slope of the clock signal at which all digital channels of the bus are ana-
lyzed.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<ClockSlope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Clock slope" on page 426
PBUS<pb>:DECTable:SHOW <DecodeTable>
If enabled, a result table is shown with decoded values and corresponding points in
time of the bus signal. Each clock edge corresponds to one row in the table.
The decode table is only available for clocked buses to check the data words.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<DecodeTable> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Show decode table" on page 426
PBUS<pb>:DECTable:COUNt?
Returns the number of rows in the decode table. Each clock edge corresponds to one
row in the table. The decode table is only available for clocked buses.
Suffix:
<pb>
.
1...4, index of the logic group
Usage: Query only
Asynchronous command
Manual operation: See "Show decode table" on page 426
Mixed signal option (option R&S MXO4-B1)