Remote control commandsMXO 4 Series
559User Manual 1335.5337.02 ─ 07
*RST: GTHan
Usage: Asynchronous command
Manual operation: See "Range" on page 161
TRIGger:EVENt<ev>:SLEW:RATE <Time>
For TRIGger:EVENt<ev>:SLEW:RANGe = INSRange and OUTRange, the slew rate
defines the center of a range which is defined by the limits "±Delta".
For TRIGger:EVENt<ev>:SLEW:RANGe = LTHan and GTHan, the slew rate defines
the maximum and minimum slew rate limits, respectively. When the signal crosses this
level, the slew rate measurement starts or stops depending on the selected slope (see
TRIGger:EVENt<ev>:SLEW:SLOPe).
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Time> Range: 1E-10 to 10000
Increment: 1E-07
*RST: 1E-10
Default unit: s
Usage: Asynchronous command
Manual operation: See "Slew rate" on page 161
TRIGger:EVENt<ev>:SLEW:SLOPe <Slope>
Sets the edge direction for the trigger.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Slope" on page 160
17.9.11 Setup & Hold
TRIGger:EVENt<ev>:SETHold:CSOurce:EDGE................................................................560
TRIGger:EVENt<ev>:SETHold:CSOurce:LEVel.................................................................560
TRIGger:EVENt<ev>:SETHold:CSOurce[:VALue]..............................................................560
TRIGger:EVENt<ev>:SETHold:HTIMe..............................................................................561
TRIGger:EVENt<ev>:SETHold:STIMe..............................................................................561
Trigger