Protocol analysisMXO 4 Series
394User Manual 1335.5337.02 ─ 07
Required options and equipment
For performing LIN decode measurements, you need the following equipment:
MXO 4 with 1 available channel. The channel can be:
Analog channels (C1-C4)
Logic channels (D0-D15) (requires option R&S MXO4-B1)
One probe
Option R&S MXO4-K520
13.6.1 LIN configuration
13.6.1.1 LIN configuration settings
Access: [Apps] key > "Protocol" tab > "LIN" > "Setup".
LIN standard
Selects the version of the LIN standard that is used in the DUT. The setting mainly
defines the checksum version used during decoding.
The most common version is LIN v.2.x. For mixed networks, or if the standard is
unknown, set the LIN standard to "Auto".
Remote command:
SBUS<sb>:LIN:STANdard on page 797
Data
Sets the source waveform of the data line.
LIN (option R&S
MXO4-K520)