Remote control commandsMXO 4 Series
539User Manual 1335.5337.02 ─ 07
Usage: Query only
Asynchronous command
Manual operation: See "Resolution in bits" on page 117
HDEFinition:STATe <State>
Enables high definition mode, which increases the numeric resolution of the waveform
signal.
Parameters:
<State> ON | OFF
ON: high definition mode
OFF: normal oscilloscope mode
*RST: OFF
Usage: Asynchronous command
Manual operation: See "State" on page 117
17.8.10 Reference clock
SENSe[:ROSCillator]:OUTPut[:ENABle] <ReferenceOutput>
Sends the internal reference clock signal to the Ref. Out connector.
Parameters:
<ReferenceOutput> ON | OFF
*RST: OFF
Usage: SCPI confirmed
Asynchronous command
Manual operation: See "Output 10 MHz ref. signal" on page 109
SENSe[:ROSCillator]:SOURce <RefSource>
Enables the use of an external 10 MHz reference signal instead of the internal refer-
ence clock.
Parameters:
<RefSource> INTernal | EXTernal
*RST: INTernal
Usage: Asynchronous command
Manual operation: See "Use external ref. clock" on page 109