Remote control commandsMXO 4 Series
643User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
17.13.3 Peak search using cursors
CURSor<cu>:FFT:SETCenter..........................................................................................643
CURSor<cu>:FFT:TOCenter........................................................................................... 643
CURSor<cu>:MAXimum[:PEAK]......................................................................................643
CURSor<cu>:MAXimum:LEFT........................................................................................ 643
CURSor<cu>:MAXimum:RIGHt....................................................................................... 644
CURSor<cu>:MAXimum:NEXT........................................................................................644
CURSor<cu>:PEXCursion.............................................................................................. 644
CURSor<cu>:THReshold................................................................................................644
CURSor<cu>:FFT:SETCenter
Sets the center frequency to the frequency value that is measured at cursor line Cu1.
Suffix:
<cu>
.
1...2, index of the cursor set
Usage: Setting only
Asynchronous command
Manual operation: See "Set center frequency to" on page 216
CURSor<cu>:FFT:TOCenter
Sets the vertical cursor line Cu1 to the center frequency.
Suffix:
<cu>
.
1...2, index of the cursor set
Usage: Setting only
Asynchronous command
Manual operation: See "Center frequency" on page 216
CURSor<cu>:MAXimum[:PEAK]
Sets both cursors to the absolute peak value.
Suffix:
<cu>
.
1...2, index of the cursor set
Usage: Event
Asynchronous command
Manual operation: See "Absolute peak" on page 216
CURSor<cu>:MAXimum:LEFT
Cursor 2 is set to the next peak to the left of the current position.