Remote control commandsMXO 4 Series
843User Manual 1335.5337.02 ā 07
SBUS<sb>:TNOS:FRAMe<fr>:TPLenth?
Returns the type length for the specified frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<TypeLenth> Range: 0 to 65535
Increment: 1
*RST: 0
Usage: Query only
Asynchronous command
SBUS<sb>:TNOS:FRAMe<fr>:TYPE?
Returns the type of the frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameType> MAC | BEACon | UNKN
*RST: MAC
Usage: Query only
Asynchronous command
17.17 Mixed signal option (option R&S MXO4-B1)
This chapter describes the remote commands of MSO option R&S MXO4-B1.
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
ā
www.rohde-schwarz.com/rc-via-scpi, chapter "Command sequence and synchroni-
zation"
Mixed signal option (option R&S
MXO4-B1)