Remote control commandsMXO 4 Series
842User Manual 1335.5337.02 ─ 07
Return values:
<FrameStart> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:TNOS:FRAMe<fr>:STATus?
Returns the state of the frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameState> OK | EPRMble | ESFD | EESD | ECRC | INComplete
EPRMble: preamble error, the hex value of the preamble field is
different from 0x55555555555555 (56 alternating 0's and 1's).
ESFD: SFD error, the hex value of the SFD field is different from
0xD5
EESD: ESD error, the value of the ESD field does not correspond
to the symbol pair "ESD, ESDOK"
ECRC: CRC error, the value of the FCS field does not match the
calculated CRC.
INComplete: the frame is incomplete
*RST: OK
Usage: Query only
Asynchronous command
SBUS<sb>:TNOS:FRAMe<fr>:STOP?
Returns the stop time of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameStop> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
Protocols