Remote control commandsMXO 4 Series
707User Manual 1335.5337.02 ─ 07
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Path> String parameter
Usage: Asynchronous command
Manual operation: See "Save as" on page 319
17.16.2 SPI (option R&S MXO4-K510)
Before MXO 4 can accept the specific bus commands, make sure that:
The decoding of the specified bus is enabled with SBUS<sb>[:STATe].
The bus is set to the correct type with SBUS<sb>:TYPE.
Configuration.........................................................................................................707
Filter...................................................................................................................... 713
Trigger...................................................................................................................719
Decode results...................................................................................................... 720
17.16.2.1 Configuration
SBUS<sb>:SPI:BORDer.................................................................................................707
SBUS<sb>:SPI:CSELect:HYSTeresis...............................................................................708
SBUS<sb>:SPI:CSELect:POLarity...................................................................................708
SBUS<sb>:SPI:CSELect:SOURce...................................................................................708
SBUS<sb>:SPI:CSELect:THReshold............................................................................... 709
SBUS<sb>:SPI:MISO:HYSTeresis................................................................................... 709
SBUS<sb>:SPI:MISO:POLarity....................................................................................... 709
SBUS<sb>:SPI:MISO:SOURce....................................................................................... 709
SBUS<sb>:SPI:MISO:THReshold....................................................................................710
SBUS<sb>:SPI:MOSI:HYSTeresis................................................................................... 710
SBUS<sb>:SPI:MOSI:POLarity....................................................................................... 710
SBUS<sb>:SPI:MOSI:SOURce........................................................................................711
SBUS<sb>:SPI:MOSI:THReshold.................................................................................... 711
SBUS<sb>:SPI:SCLK:HYSTeresis................................................................................... 711
SBUS<sb>:SPI:SCLK:SOURce....................................................................................... 711
SBUS<sb>:SPI:SCLK:THReshold....................................................................................712
SBUS<sb>:SPI:TIMeout................................................................................................. 712
SBUS<sb>:SPI:WSIZe................................................................................................... 712
SBUS<sb>:SPI:MISO:POSition....................................................................................... 713
SBUS<sb>:SPI:MISO:SCALe..........................................................................................713
SBUS<sb>:SPI:MOSI:POSition....................................................................................... 713
SBUS<sb>:SPI:MOSI:SCALe..........................................................................................713
SBUS<sb>:SPI:BORDer <BitOrder>
Selects the bit order, which determines if the data of the messages starts with MSB
(most significant bit) or LSB (least significant bit).