Remote control commandsMXO 4 Series
712User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "SCLK,MOSI,MISO,CS" on page 323
SBUS<sb>:SPI:SCLK:THReshold <Threshold>
Sets a user-defined threshold value for the line.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Threshold>
Usage: Asynchronous command
Manual operation: See "Threshold" on page 324
SBUS<sb>:SPI:TIMeout <ClockTimeout>
Sets the minimum clock idle time if a timeout on the clock line SCLK is used as limiter
between two frames.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<ClockTimeout> Range: 5E-08 to 10
Increment: 1E-06
*RST: 0.001
Default unit: s
Usage: Asynchronous command
Manual operation: See "Timeout" on page 324
SBUS<sb>:SPI:WSIZe <WordLength>
Sets the word length (or symbol size), which is the number of bits in a message. The
maximum word length is 32 bit.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<WordLength> Range: 4 to 32
Increment: 1
*RST: 8
Usage: Asynchronous command
Manual operation: See "Word length" on page 323
Protocols