Remote control commandsMXO 4 Series
711User Manual 1335.5337.02 ─ 07
SBUS<sb>:SPI:MOSI:SOURce MOSISource
Sets the input channel of the MOSI line.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
MOSISource C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9
| D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 |
R1 | R2 | R3 | R4
Usage: Asynchronous command
Manual operation: See "SCLK,MOSI,MISO,CS" on page 323
SBUS<sb>:SPI:MOSI:THReshold <Threshold>
Sets a user-defined threshold value for the line.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Threshold>
Usage: Asynchronous command
Manual operation: See "Threshold" on page 324
SBUS<sb>:SPI:SCLK:HYSTeresis <Hysteresis>
Sets a value for the hysteresis for the SCLK channel.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Hysteresis>
Usage: Asynchronous command
Manual operation: See "Threshold" on page 324
SBUS<sb>:SPI:SCLK:SOURce SCLKSource
Sets the input channel of the SCLK line.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
SCLKSource C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9
| D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 |
R1 | R2 | R3 | R4