Remote control commandsMXO 4 Series
568User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "Holdoff mode" on page 170
TRIGger:HOLDoff:MIN <RandomMinTime>
Defines the lower limit for the random time holdoff, if TRIGger:HOLDoff:MODE
TRIGger:HOLDoff:MODETRIGger:HOLDoff:MODE is set to RANDom.
Parameters:
<RandomMinTime> Range: 1E-07 to 5
Increment: 0.0002
*RST: 0.001
Default unit: s
Usage: Asynchronous command
Manual operation: See "Holdoff mode" on page 170
TRIGger:HOLDoff:TIME
<Time>
Defines the holdoff time period, if TRIGger:HOLDoff:MODE is set to TIME. The next
trigger occurs only after this time has passed.
Parameters:
<Time> Range: 1E-07 to 10
Increment: 0.0002
*RST: 0.001
Default unit: s
Usage: Asynchronous command
Manual operation: See "Holdoff mode" on page 170
17.9.15 Hysteresis
TRIGger:NOISe<m>:ABSolute........................................................................................ 568
TRIGger:NOISe<m>:EFFective?..................................................................................... 569
TRIGger:NOISe<m>:MODE............................................................................................ 569
TRIGger:NOISe<m>:PERDivision....................................................................................569
TRIGger:NOISe<m>:RELative.........................................................................................570
TRIGger:NOISe<m>[:STATe]...........................................................................................570
TRIGger:NOISe<m>:ABSolute <Absolute>
Defines a range in absolute values around the trigger level. If the signal oscillates
inside this range and thus crosses the trigger level, no trigger event occurs.
Suffix:
<m>
.
1 to 4, index of the analog channel
Trigger