Remote control commandsMXO 4 Series
569User Manual 1335.5337.02 ─ 07
Parameters:
<Absolute>
Usage: Asynchronous command
Manual operation: See "Absolute hysteresis" on page 172
TRIGger:NOISe<m>:EFFective?
Returns the hysteresis that is set by the instrument in automatic hysteresis mode.
Suffix:
<m>
.
1 to 4, index of the analog channel
Return values:
<Effective> numeric value
Usage: Query only
Asynchronous command
Manual operation: See "HW hysteresis" on page 172
TRIGger:NOISe<m>:MODE <Mode>
Selects whether the hysteresis is defined in absolute or relative values. The setting is
available only in manual hysteresis mode.
Suffix:
<m>
.
1 to 4, index of the analog channel
Parameters:
<Mode> ABS | REL
Usage: Asynchronous command
Manual operation: See "Hysteresis mode" on page 172
TRIGger:NOISe<m>:PERDivision <InDivision>
Defines a range in divisions around the trigger level in division units. If the signal oscil-
lates inside this range and crosses the trigger level thereby, no trigger event occurs.
Suffix:
<m>
.
1 to 4, index of the analog channel
Parameters:
<InDivision> Range: 0 to 5
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
Manual operation: See "Relative hysteresis" on page 172