EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 172

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TriggerMXO 4 Series
172User Manual 1335.5337.02 ─ 07
The rejection of noise by setting a hysteresis avoids unwanted trigger events caused
by noise oscillation around the trigger level.
HW hysteresis
Displays the hysteresis that is set by the instrument in automatic hysteresis mode.
Remote command:
TRIGger:NOISe<m>:EFFective? on page 569
Hysteresis mode
Selects how the hysteresis is set.
"Auto"
Automatic mode is the recommended mode. The hysteresis is set by
the instrument to reject the internal noise of the instrument.
"Manual"
The hysteresis is defined directly in absolute or relative values.
Remote command:
TRIGger:NOISe<m>:MODE on page 569
Size mode
Selects whether the hysteresis is defined in absolute or relative values. The setting is
available only in manual hysteresis mode.
Remote command:
TRIGger:NOISe<m>:MODE on page 569
Absolute hysteresis
Defines a range in absolute values around the trigger level. If the signal oscillates
inside this range and thus crosses the trigger level, no trigger event occurs.
Remote command:
TRIGger:NOISe<m>:ABSolute on page 568
Relative hysteresis
Defines a range in divisions around the trigger level in division or as percentage. If the
signal oscillates inside this range and thus crosses the trigger level, no trigger event
occurs.
Hysteresis

Table of Contents

Related product manuals