Remote control commandsMXO 4 Series
856User Manual 1335.5337.02 ─ 07
Parameters:
<DataFormat> HEX | OCT | BIN | ASCII | ASCii | SIGN | USIG
*RST: HEX
Usage: Asynchronous command
Manual operation: See "Data format" on page 425
PBUS<pb>:DATA:HEADer?
Returns the header data of the indicated bus.
For a detailed description, see DIGital<m>:DATA:HEADer?.
Suffix:
<pb>
.
1...4, index of the logic group
Usage: Query only
Asynchronous command
Manual operation: See "Data format" on page 425
PBUS<pb>:DATA[:VALues]? [<Offset>[,<Length>]]
Returns the data of the indicated logic.
Without parameters, the complete waveform is retrieved. Using the offset and length
parameters, data can be retrieved in smaller portions, which makes the command
faster. If you send only one parameter, it is interpreted as offset, and the data is
retrieved from offset to the end of the waveform.
Requirements:
PBUS<pb>:STATe is set to ON.
PBUS<pb>:DISPlay:SHBU is set to ON.
A number format is set with PBUS<pb>:DATA:FORMat.
Suffix:
<pb>
.
1...4, index of the logic group
Query parameters:
[<Offset>] Number of offset waveform points to be skipped.
Range: 0 to m. Limit: n + m <= record length
[<Length>] Number of waveform points to be retrieved.
Range: 1 to n. Limit: n + m <= record length
Return values:
<Data> List of values according to the format and content settings.
Example:
PBUS:STAT ON
PBUS:DISP:SHBU ON
PBUS:DISP:BTYP COMB
PBUS:DATA:FORMat HEX
PBUS:DATA:VAL?
Mixed signal option (option R&S MXO4-B1)