EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 852

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
852User Manual 1335.5337.02 ─ 07
PBUS<pb>:POSition <YPosition>
Sets the position of the indicated logic group waveform.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<YPosition> Range: 0 to 100
Increment: 0.1
*RST: 50
Default unit: %
Usage: Asynchronous command
PBUS<pb>:SCALe <RelativeHeight>
Sets the size of the display that is used by the indicated logic group waveform.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<RelativeHeight> Range: 3 to 100
Increment: 0.5
*RST: 10
Default unit: %
Usage: Asynchronous command
PBUS<pb>:SKEW <SkewOffset>
Sets a general delay for all digital channels.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<SkewOffset> Range: -2E-07 to 2E-07
Increment: 2E-10
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Skew" on page 428
PBUS<pb>:STATe <State>
Enables the selected logic group. The corresponding signal icon appears on the signal
bar.
Dependencies: At least one digital channel must be enabled for the selected bus, oth-
erwise the command does not work.
Mixed signal option (option R&S
MXO4-B1)

Table of Contents