Remote control commandsMXO 4 Series
852User Manual 1335.5337.02 ─ 07
PBUS<pb>:POSition <YPosition>
Sets the position of the indicated logic group waveform.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<YPosition> Range: 0 to 100
Increment: 0.1
*RST: 50
Default unit: %
Usage: Asynchronous command
PBUS<pb>:SCALe <RelativeHeight>
Sets the size of the display that is used by the indicated logic group waveform.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<RelativeHeight> Range: 3 to 100
Increment: 0.5
*RST: 10
Default unit: %
Usage: Asynchronous command
PBUS<pb>:SKEW <SkewOffset>
Sets a general delay for all digital channels.
Suffix:
<pb>
.
1...4, index of the logic group
Parameters:
<SkewOffset> Range: -2E-07 to 2E-07
Increment: 2E-10
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Skew" on page 428
PBUS<pb>:STATe <State>
Enables the selected logic group. The corresponding signal icon appears on the signal
bar.
Dependencies: At least one digital channel must be enabled for the selected bus, oth-
erwise the command does not work.
Mixed signal option (option R&S
MXO4-B1)