Remote control commandsMXO 4 Series
547User Manual 1335.5337.02 ─ 07
TRIGger:EVENt<ev>:GLITch:POLarity <Polarity>
Sets the polarity of a pulse, which is the direction of the first pulse slope.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Polarity> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Polarity" on page 150
TRIGger:EVENt<ev>:GLITch:RANGe <RangeMode>
Selects which glitches are identified: shorter or longer than the width specified using
TRIGger:EVENt<ev>:GLITch:WIDTh.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<RangeMode> SHORter | LONGer
SHORter
Glitches shorter than the specified width are identified.
LONGer
Glitches longer than the specified width are identified.
*RST: SHORter
Usage: Asynchronous command
Manual operation: See "Range" on page 150
TRIGger:EVENt<ev>:GLITch:WIDTh <Width>
Sets the length of a glitch. The instrument triggers on pulses shorter or longer than this
value, depending on the value set with TRIGger:EVENt<ev>:GLITch:RANGe.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Width> Range: 1E-10 to 10000
Increment: 0.0001
*RST: 1E-09
Default unit: s
Usage: Asynchronous command
Manual operation: See "Width" on page 150