EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 437

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Waveform generator (option R&S MXO4-B6)MXO 4 Series
437User Manual 1335.5337.02 ─ 07
Remote command:
WGENerator<wg>:VOLTage[:VPP] on page 862
Offset
Sets the vertical offset of the generated waveform.
Remote command:
WGENerator<wg>:VOLTage:OFFSet on page 861
Arbitrary Setup
Opens a menu to configure the arbitrary waveform.
See Chapter 15.1.4, "Arbitrary waveforms", on page 442.
Sweep
Opens a menu to configure the sweep.
See Chapter 15.1.3, "Sweep settings", on page 441.
Modulation
Opens a menu to configure the modulation.
See Chapter 15.1.2, "Modulation settings", on page 438.
Noise level in %
Sets the level of the noise in percentage of the set "Amplitude" output of the signal.
Remote command:
WGENerator<wg>:MODulation:NLPCent on page 868
User load
Select the user load, the load of the DUT at its connection.
You can select either a "50Ω" or a "High-Z" (high input impedance) load.
Remote command:
WGENerator<wg>:OUTPut[:LOAD] on page 859
Symmetry
Sets the symmetry of a ramp waveform, the percentage of time the waveform is rising.
By changing the symmetry of the ramp, you can create, for example, triangular wave-
forms.
50% defines symmetric triangles. Values <50% define triangles with steeper rising
edge leaned to the left. Values >50% define triangles with steeper falling edge leaned
to the right.
Remote command:
WGENerator<wg>:FUNCtion:RAMP[:SYMMetry] on page 858
Duty cycle
Sets the duty cycle for the pulse function.
The duty cycle expresses for what percentage of the period, the signal state is high.
Remote command:
WGENerator<wg>:FUNCtion[:SQUare]:DCYCle on page 859
Setup of the waveform generator

Table of Contents