Remote control commandsMXO 4 Series
799User Manual 1335.5337.02 ─ 07
TRIGger:LIN:CHKSerror <ChecksumError>
Triggers on checksum errors.
Available, if TRIGger:LIN:TYPE is set to ERRC.
Parameters:
<ChecksumError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 398
TRIGger:LIN:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, not equal, less than, less or equal than, greater than,
greater or equal than. These conditions require one data pattern
to be set with TRIGger:LIN:DMIN.
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Data setup: Condition, Pattern, Position" on page 397
TRIGger:LIN:DMIN <DataPattern>
Specifies a data pattern, or sets the start value of a data pattern range.
Parameters:
<DataPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Data setup: Condition, Pattern, Position" on page 397
TRIGger:LIN:DPOSition <DataPosition>
Sets the number of data events that are ignored, before trigger condition check of the
data starts.
Parameters:
<DataPosition> Range: 1 to 4096
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Data setup: Condition, Pattern, Position" on page 397
Protocols