Remote control commandsMXO 4 Series
553User Manual 1335.5337.02 ─ 07
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 5E-10
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "±Delta" on page 156
TRIGger:EVENt<ev>:WINDow:RANGe <RangeMode>
Selects how the signal run is compared with the window.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<RangeMode> ENTer | EXIT | WITHin | OUTSide
ENTer
Triggers when the signal crosses the upper or lower level and
thus enters the window made up of these two levels.
EXIT
Triggers when the signal leaves the window.
WITHin
Triggers if the signal stays between the upper and lower level for
a specified time. The time is defined with TRIGger:
EVENt<ev>:WINDow:TIME.
OUTSide
Triggers if the signal stays above the upper level or below the
lower level for a specified time. The time is defined with
TRIGger:EVENt<ev>:WINDow:TIME.
*RST:
ENTer
Usage: Asynchronous command
Manual operation: See "Vertical condition" on page 155
TRIGger:EVENt<ev>:WINDow:TIME <TimeRangeMode>
Available for TRIGger:EVENt<ev>:WIDTh:RANGe = WITHin and OUTSide.
Selects how the time limit of the window is defined.
You can specify the width with TRIGger:EVENt<ev>:WIDTh:WIDTh and the delta
with TRIGger:EVENt<ev>:WINDow:DELTa.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<TimeRangeMode> WITHin | OUTSide | SHORter | LONGer