EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 278

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ApplicationsMXO 4 Series
278User Manual 1335.5337.02 ─ 07
Power quality results
The results of power quality measurements are provided in two ways:
The diagram shows the graphical presentation of:
The voltage waveform
The current waveform
The power waveform that is the product of the current and voltage waveforms
The result box displays the numeric measurement results:
Voltage and current results
The voltage and current results are defined as follows:
Result Description
Current / Voltage RMS Square root of the mean of the square of the current
or voltage averaged over N cycles
Voltage Frequency Frequency of the signal
Current / Voltage Cycle Crest factor Peak value / RMS value
Power results
The power in a system is described by several physical quantities:
Power analysis (option R&S
MXO4-K31)

Table of Contents

Related product manuals