Protocol analysisMXO 4 Series
404User Manual 1335.5337.02 ─ 07
If not try adjusting the vertical and horizontal settings.
9. Set the logical thresholds:
a) Tap "Threshold setup". The threshold dialog opens.
b) For each wire, set the threshold value. A typical value is 3.5 V.
c) If necessary, set the hysteresis value.
d) Optionally, tap on "Show threshold lines".
13.6.5.2 Triggering on LIN
Prerequisites: A bus is configured for the LIN signal to be analyzed. See Chap-
ter 13.6.5.1, "Configuring LIN signals", on page 403.
1. Open "Menu" > "Trigger".
2. Set "Trigger on" to "Single event".
3. Ensure that the "Source" is set to the configured serial bus, e.g. "Serial bus 1".
4. Select the "Type".
5. Define additional settings for more complex trigger types: "Identifier", "Identifier
+Data" .
a) Tap on "Set details".
b) Set the "Identifier" condition or an identifier range.
c) Set the "Data" condition or a data range.
13.6.5.3 Exporting decode results
Once a decode has been completed, you can export the decode results of the
Prerequisites: A bus is configured for the LIN and decoded.
1. Tap on the "LIN" shortcut.
2. Tap on the "Export" tab.
3. If necessary, enable "Include details".
4. If necessary, enable "Include timing".
5. Select the "File type".
An export file is saved in the selected directory.
The LIN export file contains the following fields:
A list of all decoded frames:
Index
Start
Stop
State
LIN (option R&S
MXO4-K520)

Table of Contents