Remote control commandsMXO 4 Series
506User Manual 1335.5337.02 ─ 07
AUToscale
Performs an autoset process: analyzes the enabled channel signals, and obtains
appropriate horizontal, vertical, and trigger settings to display stable waveforms.
Rohde & Schwarz does not recommend using the autoset in remote control. To adjust
the oscilloscope remotely, especially for automated testing applications, use the
remote commands that adjust the horizontal, vertical and trigger settings.
Usage: Event
Asynchronous command
TIMebase:SCALe <TimebaseScale>
Sets the horizontal scale, the time per division, for all waveforms in the time domain,
for example, channel and math waveforms.
Parameters:
<TimebaseScale> Range: 200E-12 to 10E3
Increment: 1E-12
*RST: 20E-9
Default unit: s/div
Usage: Asynchronous command
Manual operation: See "[Scale]" on page 41
TIMebase:RANGe <TimebaseRange>
Sets the time of one acquisition, which is the time across the 10 divisions of the dia-
gram: Acquisition time = Time scale * 10 divisions.
Parameters:
<TimebaseRange> Range: Device specific to Device specific
Increment: 1E-12
*RST: Device specific
Default unit: s
Usage: Asynchronous command
Manual operation: See "Timebase range" on page 106
TIMebase:DIVisions?
Returns the number of horizontal divisions on the screen. The number cannot be
changed.
Return values:
<HorizDivCnt> *RST: 10
Usage: Query only
Asynchronous command