ACQuire:HISTory:STARt <StartAcqIndex>
Sets the index of the first (oldest) acquisition to be displayed or exported. The index is
always negative.
Parameters:
<StartAcqIndex> Range: -2147483648 to 0
Increment: 1
*RST: 0
Usage: Asynchronous command
Manual operation: See "Start acq" on page 194
ACQuire:HISTory:STOP
<StopAcqIndex>
Sets the index of the last (newest) acquisition to be displayed or exported. The newest
acquisition of the complete acquisition series always has the index "0".
Parameters:
<StopAcqIndex> Range: -2147483648 to 0
Increment: 1
*RST: 0
Usage: Asynchronous command
Manual operation: See "Stop acq" on page 194
ACQuire:HISTory:TPACq <TimePerAcq>
Sets the display time for one acquisition. The shorter the time, the faster the replay is.
Parameters:
<TimePerAcq> Range: 4E-05 to 10
Increment: 1
*RST: 0.05
Default unit: s
Usage: Asynchronous command
Manual operation: See "Display time" on page 195
ACQuire:HISTory:TSABsolute?
Returns the absolute daytime of the current acquisition (ACQuire:HISTory:
CURRent).
Return values:
<TimeAbsString> String containing the time and unit
Usage: Query only
Asynchronous command
Manual operation: See "Time stamp" on page 195