Remote control commandsMXO 4 Series
632User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "Auto Scale" on page 239
MEASurement<mg>:TRACk:OFFSet <VerticalOffset>
Sets or queries the offset of the track waveform.
If MEASurement<mg>:TRACk:CONTiunous is ON, use the command to query the
current value.
If MEASurement<mg>:TRACk:CONTiunous is OFF, the command sets the offset.
Suffix:
<mg>
.
1...16, index of the measurement
Parameters:
<VerticalOffset> Range: -1E+26 to 1E+26
Increment: 1E-06
*RST: 0
Default unit: div
Usage: Asynchronous command
Manual operation: See "Meas offset" on page 239
MEASurement<mg>:TRACk:SCALe <VerticalScale>
Sets or queries the vertical scale of the track diagram.
If MEASurement<mg>:TRACk:CONTiunous is ON, use the command to query the
current value.
If MEASurement<mg>:TRACk:CONTiunous is OFF, the command sets the scale.
Suffix:
<mg>
.
1...16, index of the measurement
Parameters:
<VerticalScale> Range: 1E-26 to 1E+26
Increment: 1E-05
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
Manual operation: See "Meas scale" on page 239
MEASurement<mg>:TRACk:DATA:HEADer?
Returns the header of track data.
Automatic measurements