EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 722

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
722User Manual 1335.5337.02 ─ 07
SBUS<sb>:SPI:FRAMe<fr>:DATA?
Returns the data words of the specified frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameData> Comma-separated sequence of integer values (N, L1, R1,…,
LN, RN). N is the number of word pairs in the frame, and {L1,R1}
…{LN,RN} are the value pairs. The values Lx and Rx are associ-
ated with the MOSI and the MISO channel, respectively. If a
channel is disabled, an empty value is returned.
Usage: Query only
Asynchronous command
SBUS<sb>:SPI:FRAMe<fr>:STARt?
Returns the start time of the specified frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameStart> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:SPI:FRAMe<fr>:STATus?
Returns the overall state of the specified frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameState> OK | VOID | INComplete | LENGth
OK: the frame is valid.
VOID: the frame is empty.
INComplete: INComplete word. The word does not have the
expected word length.
LENGth: The frame is not contained in the acquisition.
*RST: OK
Protocols

Table of Contents

Related product manuals