Remote control commandsMXO 4 Series
875User Manual 1335.5337.02 ─ 07
Parameters:
<CoupleEnable> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Generator enable" on page 446
WGENerator<wg>:COUPling:AMPLitude <CplAmplitude>
Enables the coupling of all amplitude parameters of the generators.
Suffix:
<wg>
.
1..2
Specifies the prime generator.
Parameters:
<CplAmplitude> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Frequency parameters" on page 446
WGENerator<wg>:COUPling:PHASeshift <PhaseShift>
Sets the phase shift between the waveform of Gen1 and Gen2 when the frequency
parameters of the two waveforms are coupled.
Suffix:
<wg>
.
1..2
Specifies the prime generator.
Parameters:
<PhaseShift> Range: -180 to 180
Increment: 1
*RST: 0
Default unit: °
Usage: Asynchronous command
Manual operation: See "Phase Gen1 - Gen2" on page 447
WGENerator<wg>:COUPling[:FREQuency] <CplFreq>
Enables the coupling of all frequency parameters of the generators.
Suffix:
<wg>
.
1..2
Specifies the prime generator.
Parameters:
<CplFreq> ON | OFF
*RST: OFF
Waveform generator (option R&S MXO4-B6)