Instrument Function
R&S
®
SMU200A
134Operating Manual 1007.9845.32 ─ 15
Current Sensors
Section "Current Sensors" lists the sensors currently connected to the generator and
indicates their serial number, the revision state and some features.
SCPI command:
n.a.
Update
Section "Update" provides access to the file system in order to select a file for an R&S
NRP sensor update (Button "Select NRP File"), the selected file is indicated to the left of
the button. On the left side, the sensor to be updated is selected.
Button "Run Update" starts the update. If the update is interrupted for example by acci-
dently pulling off the sensor, selection "Rescue" is offered to restart the update process.
Prerequisite is that no other sensor is connected to the instrument.
The procedure is as follows:
●
Do not reconnect the sensor but keep is ready to be connected
●
Select "Rescue" in the left sensor selection field
●
Activate "Run Update"
●
Confirm query in message box
●
Connect sensor within 4 seconds
The update starts, a bar informs about the progress.
SCPI command:
SENSe<ch>[:POWer]:TYPE on page 565
SENSe<ch>[:POWer]:SVERsion on page 564
SENSe<ch>[:POWer]:SNUMber on page 563
5.2.3.7 Update
After a firmware update it is occasionally required to also update the "PCI-FPGA". This
is enabled in the "Update" dialog.
At the first start of the new firmware, a message appears during the boot process if a
"PCI-FPGA" update is required. Execute a PCI-FPGA update by pressing the "PCI-
FPGA" button.
General Instrument Settings