EasyManua.ls Logo

Rohde & Schwarz SMU200A - Bert:result

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
514Operating Manual 1007.9845.32 ─ 15
:BERT:RESult?
Queries the result of the last BER measurement/BLER measurement. The response
consists of seven results separated by commas. In the first measurement following the
start, intermediate results for the number of data bits/data blocks, error bits/errors and
error rate are also queried. In the following measurements (only for BERT:SEQ AUTO/
for BLER:SEQ AUTO) only the final results of each single measurement are queried
Note: At the end of a measurement, the restart of a new one is delayed until the first
measurement result has been queried with BERT:RES. The resulting brief measurement
interruption is irrelevant because the subsequent measurement will be synchronized
within 24 data bits.