Manual Operation
R&S
®
SMU200A
93Operating Manual 1007.9845.32 ─ 15
This button is available only if the history of the messages is displayed.
SCPI command:
SYSTem:ERRor:ALL on page 791
Each time a SYST:ERR:ALL? query is sent, the error queue is returned and at the same
time cleared.
History
Calls the list of all messages that have occurred since instrument switch-on. The most
recent messages are displayed at the top of the list. When the button is pressed again,
the list of current messages is displayed.
SCPI command:
SYSTem:ERRor[:NEXT] on page 793 or STATus:QUEue[:NEXT] on page 788
Each time a SYST:ERR? or STAT:QUE? query is sent, the oldest entry in the error queue
is returned and at the same time cleared in the list.
4.2.4 Block Diagram
The block diagram shows provided options, signal configuration and the currently
selected signal flow of the generator with inputs and outputs used. Signal generation can
be completely operated from the block diagram. The highlighted function block can be
directly switched on and off with the TOGGLE ON/OFF (CTRL+T) key. Pressing the Enter
opens the associated setting menu.
4.2.4.1 Function Blocks in the Block Diagram
Each block represents a function of signal generation. The function is indicated in the
headline of the block, e.g. "Baseband". In this block, the digital modulation signal, the
digital standards, arbitrary waveform generation and multicarrier CW are set, for instance.
In the check box, the respective function can be quickly activated/ deactivated with the
TOGGLE ON/OFF (CTRL+T) key. After activation, the block is displayed in blue.
Display