Remote Control Commands
R&S
®
SMU200A
524Operating Manual 1007.9845.32 ─ 15
CAL:ALL:MEAS?
starts the adjustment of all functions for the entire instrument.
Response: "0"
adjustment has been performed successfully.
Example:
Query only
SCPI conform
Usage:
:CALibration<hw>:BBIN[:MEASure]?
The command starts adjustment of the analog I/Q input. The I/Q input is adjusted with
respect to DC offset and gain.
Return values: 
0|1<Measure>
CAL:BBIN:MEAS?
starts the adjustment of the analog I/Q input.
Response: "0"
adjustment has been performed successfully.
Example:
Query only
SCPI conform
Usage:
:CALibration<hw>:FMOFfset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
Return values: 
0|1<Measure>
CAL:FMOF?
starts the adjustments for the FM/Phim modulator.
Response: "0"
the adjustments have been performed successfully
Example:
Query onlyUsage:
:CALibration<hw>:FREQuency[:MEASure]?
The command starts all adjustments which affect the frequency.
In case of two-path instruments, this affects the selected path.
Return values: 
0|1<Measure>
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Example:
Query onlyUsage:
CALibration Subsystem