Remote Control Commands
R&S
®
SMU200A
773Operating Manual 1007.9845.32 ─ 15
SCPI conformUsage:
[:SOURce<hw>]:PULM:STATe <State>
The command activates/deactivates the pulse modulation.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
PULM:STAT ON
activates pulse modulation with the modulation source selected
under SOURce:PULM:SOURce.
Example:
SCPI conformUsage:
7.15.19
SOURce:ROSCillator Subsystem
This subsystem contains the commands for setting the external and internal reference
frequency.
The commands of this subsystem are not affected by an instrument reset (*RST
on page 492).
[:SOURce]:ROSCillator:EXTernal:FREQuency..................................................................773
[:SOURce]:ROSCillator:EXTernal:RFOFf:STATe...............................................................774
[:SOURce]:ROSCillator:EXTernal:SBANdwidth.................................................................774
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue..............................................................774
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]............................................................775
[:SOURce]:ROSCillator:SOURce.....................................................................................775
[:SOURce]:ROSCillator:EXTernal:FREQuency <Frequency>
The command informs the instrument of the frequency of the external reference.
Parameters:
5MHZ|10MHZ|13MHZ
Hz
<Frequency>
Default unit:
ROSC:SOUR EXT
selects the external source. The reference must be input at the
REF IN input.
ROSC:EXT:FREQ 5MHz
informs the instrument that the external reference frequency is
5MHz.
Example:
Example:
SCPI conformUsage:
SOURce Subsystem