Instrument Function
R&S
®
SMU200A
144Operating Manual 1007.9845.32 ─ 15
5.2.3.13 Remote Channel Settings
The "Remote Channel Settings" dialog provides access to the settings for remote control.
The "Remote Channel Settings" dialog is opened using the SETUP or MENU key under
"Remote".
GPIB channel address
Sets the GPIB address of the instrument.
SCPI command:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess on page 789
Visa Resource Strings
Indicates the visa resource strings, used for remote control of the instrument. A separate
string is provided for remote control via the different interfaces.
SCPI command:
:SYSTem:COMMunicate:NETWork:RESource on page 790
:SYSTem:COMMunicate:GPIB:RESource on page 790
Local (Remote Channel Settings)
Switches the instrument to operate in local control mode.
Switching from remote to local control mode can be also done with one of the following
actions:
●
manually with the LOCAL key on the front panel
●
with the interface command >L via the remote-control interface
●
with the key combination CTRL + Q.
SCPI command:
>L
5.2.3.14 LXI Status
The "LXI - LAN eXtensions for Instruments - Status..." dialog displays the settings and
status of the LAN and allows to reset the LAN connection.
For more information about LXI, see chapter 2.9, "LXI Configuration", on page 58.
General Instrument Settings