Remote Control Commands
R&S
®
SMU200A
526Operating Manual 1007.9845.32 ─ 15
:CALibration<hw>:LEVel:STATe <State>
The command switches on or off internal level correction.
Parameters:
ON|OFF
ON
<State>
*RST:
CAL:LEV:STAT
switches on Level correction.
Example:
:CALibration<hw>:LEVel:HACCuracy[:STATe] <State>
Enables/disables level adjustment with high accuracy.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
CAL:LEV:HACC:STAT ON
switches on Hight Accuracy Level Adjustment
Example:
7.6 CLOCk Subsystem
The CLOCk system contains the commands for configuration of the signals at the clock
output and input connectors.
:CLOCk:INPut:FREQuency.............................................................................................526
:CLOCk:INPut:SLOPe....................................................................................................526
:CLOCk:OUTPut:MODE..................................................................................................527
:CLOCk:OUTPut:SOURce..............................................................................................527
:CLOCk:INPut:FREQuency?
The command sets the measured frequency of the external clock signal. An external clock
reference must be supplied at the CLOCK input.
Return values:
float<Frequency>
CLOC:INP:FREQ
queries the measured frequency of the external clock reference.
Example:
Query onlyUsage:
:CLOCk:INPut:SLOPe <Slope>
The command sets the active slope of an externally applied clock signal at the CLOCK
connector.
CLOCk Subsystem