Remote Control Commands
R&S
®
SMU200A
516Operating Manual 1007.9845.32 ─ 15
:BLER:SEQuence <Sequence>
:BERT:SEQuence <Sequence>
Selects the type of measurement: single or continuous measurement.
These commands are automatically set to AUTO by BERT:STARt or BLER:STARt.
Parameters:
AUTO|SINGle
AUTO
Continuous measurement. If one or both termination criteria are
met, the measurement in progress is terminated. At the end of a
measurement, the restart of a new one is delayed until the first
measurement result has been queried with BERT:RES. The
resulting brief measurement interruption is irrelevant because the
subsequent measurement will be synchronized within 24 data bits.
SIINGle
Single measurement. A single measurement must be started
with :TRIG:BERT. A single measurement is terminated once the
set number of data bits/blocks or number of errors is reached
AUTO
<Sequence>
Default unit:
BERT:SEQ SING
selects single measurement.
TRIG: BERT
starts the single measurement.
Example:
Example:
Dependencies: These commands are automatically set to AUTO
by BERT:STARt or BLER:STARt
Options:
:BLER:SETup:CLOCk[:POLarity] <Polarity>
:BERT:SETup:CLOCk[:POLarity] <Polarity>
Sets the polarity of the active slope of the feedback clock.
Parameters:
RISing|FALLing
RISing
The positive slope of the clock signal is active.
FALLing
The negative slope of the clock signal is active.
RISing
<Polarity>
Default unit:
BERT:SET:CLOC FALL
selects the falling slope of the clock signal as the active slope.
Example:
:BLER:SETup:CORDer <Corder>
Sets the byte order of the checksum (CRC).
BERT and BLER Subsystems