Remote Control Commands
R&S
®
SMU200A
523Operating Manual 1007.9845.32 ─ 15
Parameters:
float
0.01 to 2.0
1.00V
Volt
<Threshold>
Range:
*RST:
Default unit:
INP:BERT:THR 1 V
A high/low threshold of 1 volt is set at the BERT inputs. The signal
is high (active) for a signal voltage of 1 volt and higher.
Example:
:TRIGger:BLER[:IMMediate]
:TRIGger:BERT[:IMMediate]
Triggers a single bit error rate or block error rate measurement if the single trigger mode
(BERT|BLER:SEQ SING) is selected.
BERT:SEQ SING
selects the single trigger mode.
TRIG:BERT
starts a single bit error rate measurement.
Example:
Example:
EventUsage:
7.5 CALibration Subsystem
The CALibration system contains the commands for adjustment. Adjustment is triggered
by the query commands. The response "0" indicates error-free adjustment, and the
response "1" means that an error occurred during adjustment.
:CALibration:ALL[:MEASure]...........................................................................................523
:CALibration<hw>:BBIN[:MEASure].................................................................................524
:CALibration<hw>:FMOFfset[:MEASure]...........................................................................524
:CALibration<hw>:FREQuency[:MEASure].......................................................................524
:CALibration<hw>:IQModulator:FULL...............................................................................525
:CALibration<hw>:IQModulator:LOCal..............................................................................525
:CALibration<hw>:LEVel[:MEASure]................................................................................525
:CALibration<hw>:LEVel:STATe......................................................................................526
:CALibration<hw>:LEVel:HACCuracy[:STATe]..................................................................526
:CALibration:ALL[:MEASure]?
The command starts all internal adjustments for which no external measuring equipment
is needed.
With two-path instruments, adjustment is performed for both paths.
Return values:
0|1<Measure>
CALibration Subsystem