Remote Control Commands
R&S
®
SMU200A
500Operating Manual 1007.9845.32 ─ 15
{DATA BITLENGTH: 444}
BB:DM:DLIS:TAG? 'DATA BITLENGTH'
queries the content of the DATA BITLENGTH tag of the selected
data list file.
Response: '444'
the data list is 444 bits long.
Example:
Setting onlyUsage:
{DATA LIST-Length: #d0d1...dx...dN-1…}
(mandatory for data lists)
The tag contains the actual bit sequence of the data list in binary format.
Setting parameters:
Defines the number of bytes in the DATA LIST tag in ASCII
Format (for format see WAVEFORM length).
Data bits in binary format (8-bit unsigned characters, MSB first).
Length
dx
DATA LIST-17: #d0d1...dx...d127}
16 bytes containing 128 data bits, first bit is the MS bit of the first
byte.
Example:
Setting onlyUsage:
{DATE: yyyy-mm-dd;hh:mm:ss}
(optional)
The tag contains the date and time at which the file was created. The year must be
expressed as four digits. The instrument does not analyze this tag.
{DATE: 2009-04-02;14:32:12}
BB:ARB:WAV:TAG? 'DATE'
queries the content of the DATE tag of the selected waveform file.
Response: '2009-04-02;14:32:12'
the waveform was created on April 2, 2009 at 14 hrs 32 min
Example:
Setting onlyUsage:
{CONTROL LENGTH: ControlLength}
(optional / recommended for marker and control lists)
The tag contains the length of the control list in ASCII format. If this tag is not used, the
marker and control list length is determined by the highest position used in one of the
LIST tags. It is recommended to set this value to be equal to the number of samples in
order to keep marker signals and wave data synchronized.
R&S Signal Generator Waveform and List Format