Remote Control Commands
R&S
®
SMU200A
563Operating Manual 1007.9845.32 ─ 15
SENS:SOUR USER
selects user-defined source.
SENS:FREQ 2.44 GHz
enters the RF frequency of the source which is 2.44 GHz.
Example:
SENSe<ch>[:POWer]:OFFSet <Offset>
The command enters a level offset which is added to the measured level value after
activation with command SENSe[:POWer]:OFFSet:STATe ON. This allows e.g. an
attenuator in the signal path to be taken into account.
Parameters:
float
-100.0 to 100.0
0
dB
<Offset>
Range:
*RST:
Default unit:
SENS:POW:OFFS 10.0
sets a level offset of 10 dB
Example:
SENSe<ch>[:POWer]:OFFSet:STATe <State>
The command activates the addition of the level offset to the measured value. The level
offset value is set with command SENSe[:POWer]:OFFSet.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
SENS1:POW:OFFS 0.4dB
sets a level offset of 0.4 dB
SENS1:POW:OFFS:STAT ON
a level offset of 0.4 dB is added to the measured value.
Example:
SENSe<ch>[:POWer]:SNUMber?
The command queries the serial number of the sensor.
Return values:
string<Snumber>
SENS:SNUM?
queries the serial number.
Example:
Query onlyUsage:
SENSe, READ and INITiate Subsystems