Loading...
Instrument Function
R&S
®
SMU200A
160Operating Manual 1007.9845.32 ─ 15
File List
Selects the file with the desired instrument configuration.
SCPI command:
:MMEMory:CATalog on page 545
Exclude Frequency
The current frequency is retained when a stored instrument setting is loaded.
SCPI command:
[:SOURce<hw>]:FREQuency[:CW|:FIXed]:RCL on page 713
Exclude Level
The current level is retained when a stored instrument setting is loaded.
SCPI command:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL on page 766
Recall
Load the selected configuration.
If an instrument setting in which a sweep was activated is stored, the sweep is started
when the recall command is called.
If an instrument setting which accesses lists is stored, this list is also loaded.
If the list has been deleted in the meantime, an error message appears when the instru-
ment setting is loaded. If the list has been overwritten in the meantime, the new entries
will be used.
SCPI command:
:MMEMory:LOAD:STATe on page 549
*RCL on page 492
Recall Immediate x
Loads the selected configuration from one of the three intermediate memories.
If an instrument setting in which a sweep was activated is stored, the sweep is started
when the recall command is called.
If an instrument setting which accesses lists is stored, this list is also loaded.
If the list has been deleted in the meantime, an error message appears when the instru-
ment setting is loaded. If the list has been overwritten in the meantime, the new entries
will be used.
A message appears if no instrument configuration is stored in this memory.
SCPI command:
*RCL on page 492
5.2.8.4 File Manager
The "File Manager" dialog provides all the functions required for file management. Direc-
tories can be created, and files copied, deleted and moved between the directories on
the drives (hard disk, memory stick and network drives).